Routine Maintenance and Cleaning
Routine Maintenance
The Odyssey DLx requires only minimal maintenance. However, as with any equipment utilizing electrical voltages, there is a danger of fire or electrical shock if the equipment is not properly maintained.
LICORbio recommends that you routinely inspect the system and the scanning surface. The following are some general maintenance guidelines:
- Wipe all chemical spills from the case and/or scanning surface to prevent damage to the surface coating.
- Inspect all cables and power cords for evidence of fraying, exposed wire, or loose connections.
- Keep the scanning surface free of organic solvents and other combustibles.
- Clean the exterior case parts with warm water and a damp cloth. The exterior case parts are painted with a durable urethane coating that is resistant to chemical spills. Do not use scouring compounds or solvents (e.g., acetone, benzene, carbon tetrachlorides, lacquer thinner, or alcohol) to clean the case.
Cleaning the Scanning Surface
It is very important that the scanning surface glass and silicone mat be free of smudges, dust, and dye before scanning any samples on the Odyssey DLx. Contaminated surfaces in contact with the membrane surface may cause blotches and streaks that cannot be removed with further washing.
- Thoroughly clean the glass scanning surface with ultrapure water and a lint-free wipe.
- Repeat the wash with 70% ethanol and a lint-free wipe or 70% isopropanol and a lint-free wipe. If additional cleaning is necessary, 100% methanol may be used.
- If the silicone mat is used, rinse it under warm water. Gentle lab soap may be used, but the soap must be completely rinsed away before use. The silicone mat may also be rinsed with isopropanol or ethanol if needed. Dry the silicone mat with a lint-free tissue.